Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
Page 48 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
9.3
EIA-485 Connector
To connect a controller, a computer or a bus system over a longer distance, an EIA-
485 interface is provided.
You can either establish a point-to-point connection with a control unit or you can
access several devices via a bus structure.
The IGM11 always operates as slave unit at this bus. This means it will transmit data
only after a request.
Figure 22 EIA-485 Bus Structure
9.3.1
EIA-485 Bus Terminal
The bus line must be terminated with a resistance network on both sides.
For this purpose, a resistance of 150 Ω is switched on the first and last device
between the data lines RS485P and RS485N.
With the IGM11 the bus terminal resistance can be mounted in the connector casing.