Gaussmeter IGM11 Operating Instructions
Chapter 10 Table of Error Messages
Page 96 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
203
EIA232 buffer overflow
Buffer overflow on the serial interface EIA-232.
If necessary, reduce the bit rate. Check the
command.
300
EIA485 parity error
Character parity error on the serial interface
EIA-485.
Check the settings of the external controller
and of the device. Possibly, the transmission
path is disturbed.
301
EIA485 framing error
Character framing error on the serial interface
EIA-485.
Check the settings of the external controller
and of the device. Possibly, the transmission
path is disturbed.
302
EIA485 overrun error
Character overrun error on the serial interface
EIA-485.
Possibly, reduce the bit rate.
303
EIA485 buffer overflow
Buffer overflow on the serial interface EIA-485.
Possibly, reduce the bit rate. Check the
command.
304
EIA485 checksum error
BCC character error on the serial interface.
EIA485.
Check the settings of the external controller
and of the device. Possibly, the transmission
path is disturbed.
305
EIA485 adress error
Error in the address structure on the serial
interface EIA485.
Check the settings of the external controller
and of the device. Possibly, the transmission
path is disturbed.
nnn
Undefined error
Undefined error.
Forward the error number to the service.