Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
Page 84 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
9.12.5
Memory Functions
9.12.5.1
:PAR:SAVE
Description:
Save parameters, units and ranges in non-volatile memory.
Mode:
Command
Parameter:
none
Response:
none
*RST
Value:
not relevant
Example:
send
:PAR:SAVE<LF>
9.12.5.2
:PAR:SAVW?
Description:
Save parameters, units and ranges in non-volatile memory. After
that, the parameter status will be returned.
Mode:
Query
Parameter:
none
Response:
0|1
1
: Storage error
*RST
Value:
not relevant
Example:
send
:PAR:SAVW<LF>
receive 0<CR><LF>
9.12.5.3
:PAR:DEFA
Description:
Parameters, units and ranges are reset to default.
Mode:
Command
Parameter:
none
Response:
none
*RST
Value:
not relevant
Example:
send
:PAR:DEFA<LF>
Trial Version
Subje
ct to chan
ge