Industrial Gaussmeter IGM11 Operating Instructions
Chapter 10 Table of Error Messages
Page 95 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
10
Table of Error Messages
Errors inside the device are stored in the error queue. It can be read out via an SCPI
command or via the set-up menu.
The messages have the following meanings:
Error
Number
Error Text
Reason / Solution
0
No error
There is no error.
-101
Invalid character
SCPI error: An invalid character has been
detected. Check the command.
-102
Syntax error
SCPI error: An invalid parameter has been
detected. Check the command.
-113
Undefined header
SCPI error: An invalid command has been
detected.
Check the command.
-121
Invalid character in number
SCPI error: An invalid character in a number
has been detected. Check the command.
-151
Invalid string data
SCPI error: An invalid character in the text has
been detected. Check the command.
-222
Data out of range
SCPI error: Numerical value exceeds the
allowed limits.
Check the command.
-224
Illegal parameter value
SCPI error: Invalid parameter has been
detected. Check the command.
-350
Too many errors
Error queue is full.
522
Output buffer overflow
Output buffer for interface is too small. Check
the command.
100
Param. memory error
Parameter memory data are inconsistent.
Check the parameter settings.
101
Probe parameter error
Calibration data of the measuring probe
cannot be queried by the device.
Re-connect the probe or try another probe.
102
Internal parameter error
Internal calibration data are inconsistent.
Switch the device off and on.
103
Internal error WD
104
Internal error ADRR
105
Internal error STCK
An internal error has occurred.
Switch the device off and on.
200
EIA232 parity error
Character parity error on the serial interface
EIA-232.
Check the settings of the external controller
and of the device. Possibly, the transmission
path is disturbed.
201
EIA232 framing error
Character frame error on the serial interface
EIA-232.
Check the settings of your controller and of the
device. Possibly, the transmission path is
disturbed.
202
EIA232 overrun error
Character overrun error on the serial interface
EIA-232.
If necessary, reduce the bit rate.
Trial Version