Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
Page 76 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
9.12.2
Main Commands
9.12.2.1
:MODE[?]
Description:
Set or query the DC or AC field operating mode.
Mode:
Command and query
Parameter:
{DC|AC|PEAK}
DC
DC operation mode
AC
AC operation mode
PEAK
DC field peak value detection
Response:
Operating mode
*RST
Value:
DC
Example:
send
:MODE AC<LF>
send
:MODE?<LF>
receive AC<CR><LF>
9.12.2.2
:NULL
Description:
Null compensation of the measuring probe. This function should
only be carried out in sufficiently field-free areas.
Mode:
Command
Parameter:
none
Response:
none
*RST
Value:
not relevant
Example:
send
:NULL<LF>
Trial Version
Subje
ct to chan
ge