Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
Page 58 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
9.7.2
Operating Mode SHORT
In this operating mode simple one-character commands are evaluated and replied
depending on the command type. The IGM11 will only reply on request. The
command line must start with <STX> and terminate with <ETX>. The reply will also
be enclosed between <STX> and <ETX>. There is no distinction between upper and
lower case.
Command
Description
Reply
Reply format
<STX>?<ETX>
Read out current measuring
value
Current measuring value
s = +|-
n = 0..9
<STX>snnnn.<ETX>
<STX>snnn.n<ETX>
<STX>snn.nn<ETX>
<STX>sn.nnn<ETX>
<STX>R?<ETX>
Read out measuring range
Current measuring range
n = 0..3
<STX>b<ETX>
<STX>R0<ETX>
Set measuring range to the
most sensitive range (e.g.
10 mT)
None
--
<STX>R1<ETX>
Analogous to R0, second
most sensitive range (e.g.
100 mT)
None
--
<STX>R2<ETX>
Analogous to R0, third most
sensitive range (e.g.
1000 mT)
None
--
<STX>R3<ETX>
Set measuring range to least
sensitive measuring range
(e.g. 4500 mT)
None
--
<STX>O?<ETX>
Read out signal outputs
Status of the signal outputs
One character per output:
0 (0x30h) or 1 (0x31h)
<STX>oooo<ETX>
<STX>N<ETX>
Set probe to zero
None
--
<STX>P<ETX>
Set peak to zero
None
--
<STX>S?<ETX>
Read out status
Current status
s = S|R
S: Stop; R: Run
<STX>s<ETX>