Industrial Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
Page 77 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
9.12.2.3
:RANGe:SET
Description:
Set the measuring range of the gaussmeter.
Mode:
Command
Parameter:
{0|1|2|3}
0
Most sensitive range
3
Least sensitive range
Response:
none
*RST
Value:
3
Example:
send
:RANG:SET 2<LF>
Meaning of the parameters with different units and measuring modes.
Parameter
Tesla
Gauss
Oersted
A/m
3
4.5 / 10T
45 / 100 kG
45 / 100 kOe
3800 / 8500 kA/m
2
1 T
10 kG
10 kOe
1000 kA/m
1
100 mT
1 kG
1 kOe
100 kA/m
DC field measurement
and
Fast peak value
measurement
0
10 mT
100 G
100 Oe
10 kA/m
3
3.0 / 6.5 T
30 / 65 kG
30 / 65 kOe
2500 / 5500 kA/m
2
1 T
10 kG
10 kOe
1000 kA/m
1
100 mT
1 kG
1 kOe
100 kA/m
AC field measurement
0
10 mT
100 G
100 Oe
10 kA/m
9.12.2.4
:RANGe?
Description:
Query the current measuring range.
Mode:
Query
Parameter:
None
Response:
{0|1|2|3}
0
Most sensitive range
3
Least sensitive range
*RST
Value:
not relevant
Example:
send
:RANG?<LF>
receive 3<CR><LF>