Industrial Gaussmeter IGM11 Operating Instructions
Chapter 12 Technical Data
Page 101 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
Units
Tesla
Gauss
Oersted
Ampere/meter
4.5 T
[10T]
(1 mT)
45 kG
[100 kG]
(10 G)
45 kOe
[100 kOe]
(10 Oe)
3800 kA/m
[8500 kA/m]
(1 kA/m)
1 T
(100 µT)
10 kG
(1 G)
10 kOe
(1 Oe)
1000 kA/m
(100 A/m)
100 mT
(10 µT)
1 kG
(100 mG)
1 kOe
(100 mOe)
100 kA/m
(10 A/m)
Meas. ranges
DC-field measurement
(resolution)
[with high-field probe]
10 mT
(1 µT)
100 G
(10 mG)
100 Oe
(10 mOe)
10 kA/m
(1 A/m)
3.0 T
[6,5 T]
(10 mT)
30 kG
[65 kG]
(100 G)
30 kOe
[65 kOe]
(100 Oe)
2500 kA/m
[5500 kA/m]
(10 kA/m)
1 T
(1 mT)
10 kG
(10 G)
10 kOe
(10 Oe)
1000 kA/m
(1 kA/m)
100 mT
(100 µT)
1 kG
(1 G)
1 kOe
(1 Oe)
100 kA/m
(100 A/m)
Meas. ranges
AC-field measurement
(resolution)
[with high-field probe]
10 mT
(10 µT)
100 G
(100 mG)
100 Oe
(100 Oe)
10 kA/m
(10 A/m)
4.5 T
[10T]
(10 mT)
45 kG
[100 kG]
(100 G)
45 kOe
[100 kOe]
(100 Oe)
3800 kA/m
[8500 kA/m]
(10 kA/m)
1 T
(1 mT)
10 kG
(10 G)
10 kOe
(10 Oe)
1000 kA/m
(1 kA/m)
100 mT
(100 µT)
1 kG
(1 G)
1 kOe
(1 Oe)
100 kA/m
(100 A/m)
Meas. ranges
Fast pulse meas.
(resolution)
[with high-field probe]
10 mT
(10 µT)
100 G
(100 mG)
100 Oe
(100 mOe)
10 kA/m
(10 A/m)
Peak hold t
signal
> 1 ms
Standard transversal- and axial probes; see data sheets
High-field transversal- and axial probes; see data sheets
(Special design on request)
All probes of the HGM09 series via adapter cable
Probes
All probes:
Active range
⌀
0.15 mm
Integrated parameter memory
Environmental conditions
Operating conditions Specified accuracy for 0 °C to 40 °C
Relative operating humidity Up to 80 % relative humidity at temperatures to 30 °C,
linear decreasing to 50 % relative humidity at 40 °C
Altitude 0 – 2000 metere according to IEC 61010-1 2nd Edition CAT III, 1000 V
Degree of Pollution Pollution degree II
Notes: The technical data apply to a one-hour warm-up phase.