Industrial Gaussmeter IGM11 Operating Instructions
Chapter 8 Signal Interfaces
Page 43 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
8.2
Wiring of the Inputs
Option1
: The external control switches the external positive potential.
Figure 15 Signal Input Plus Switched
This option is preferably used in European countries in order to connect a PLC
control.
Option
2
: The control provided by the customer switches the external negative
potential.
Figure 16 Signal Input Minus Switched
This option is preferably used in America and Japan in order to connect a PLC
control.
Insulated DC input
Used as current sinking inputs
Insulated DC input
Used as current sourcing inputs
Special features
Do not connect!
Special features.
Do not connect!