Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
Page 70 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
9.11.6
Memory Functions
:PAR:SAVE
Store the set parameters
:PAR:SAVW?
Store the set parameters and queries the status
:PAR:DEFA
Default parameter settings
9.11.7
Serial Interfaces
:SERI:COMA:BAUD[?]
{1200|2400|4800|9600|14K
4|19K2|38K4}
Bit rate of EIA-232
:SERI:COMA:STOP[?] {1|2}
Stop bits of EIA-232
:SERI:COMA:PARY[?]
{NONE|EVEN|ODD}
Parity of the EIA-232 interface
:SERI:COMA:MODE[?]
{SCPI|SHOR|FLOW}
Operating mode of the EIA-232 interface
:SERI:COMB:BAUD[?]
{1200|2400|4800|9600|14K
4|19K2|38K4}
Bit rate of EIA-485
:SERI:COMB:MODE[?]
{BUS|P2P}
Operating mode for the EIA-485 interface
:SERI:COMB:ADR[?]
{0..31}
Bus address of the EIA-485 interface
:IP:IP1[?] .. :IP:IP4[?]
{0..255}
IP address of the LAN interface
:IP:MAS1[?] .. :IP:MAS4[?]
{0..255}
IP subnet mask of the LAN interface
9.11.8
Digital Interfaces
:DGIO:IN1? .. :DGIO:IN3?
Query the invidual digital inputs
:DGIO:IN?
Query all digital inputs
:
DGIO:SET:IN{x}:FUNC[?]
{
NO|HOLD|NLPK|NULL
}
Function of the single digital inputs
x = 1..3
:
DGIO:SET:IN{x}:LOG[?]
{
0|1
}
Logic of the single digital inputs
x = 1..3
:DGIO:SET:OUT{x}:FUNC[?]
{
NO|IFHI|IFLO|IFIN|IFOU|
IFOV|IFRN
}
Function of the single digital outputs
x = 1..4
:
DGIO:SET:OUT{x}:LOG[?]
{
0|1
}
Logic of the single digital outputs
x = 1..4
:DGIO:SET:OUT{x}:LOW[?]
Lower limit value for the compare function of
the single digital outputs
x = 1..4
:DGIO:SET:OUT{x}:LOW[?]
Lower limit value for the compare function of
the single digital outputs
x = 1..4
:DGIO:SET:OUT{x}:HIGH[?]
Upper limit value for the compare function of
the single digital outputs
x = 1..4
Trial Version
Subje
ct to chan
ge