Note on technical drawings
In general, all specifications are in
millimetres [mm]
, unless otherwise or directly stated in the technical
drawings.
2
Safety and general information
2.1
Safety
Read before installing, commissioning, maintaining, or operating the product.
Take note of, and observe all safety and operating instructions, to avoid malfunction or damage to the product.
SCU 835
:
ESD Precautions
To avoid damage to the control unit and to the sensor cable, disconnect power before adding or removing
cables or printed circuit boards.
When working on an open unit provisions against
ESD
must
be applied, such as an earthing mat with
wrist strap.
Connections may only be made while the unit is
switched off
.
Appropriate shipping, handling, and storage, as well as professional installation, commissioning and operation are
all required to assure safe functioning of the equipment. Never pour any liquid into the housing of the equipment.
Short circuits and fire could result. Do not drop the equipment.
Operate the device only at temperatures between
-25 °C
and
+70 °C
. Avoid exposure to moisture
Get the equipment checked by manufacturer, if:
moisture has penetrated into the equipment,
it does not work well, or it does not work according to the user's manual,
it was dropped or damaged,
it shows obvious signs of breakage.
Sensor cable:
Attention
The following precautions must always be adhered to in order to not damage the sensors inside the
cable. Failing to comply with these precautions voids warranty.
Humidity
The sensor cable must always be protected from humidity, i.e. the cable ends
must
be sealed while sto-
ring or mounting the cable.
Bending
The minimum bending radius of the sensor cable is
25 cm
. This limit must be adhered to at all times in
order to not damage the sensors inside the cable!
Installation temperature
The sensor cable may
not
be installed below
+10 °C
of its own temperature (
+ 50 °F
)! The sensor cable
must have an own temperature between
+10 °C ... +40 °C
during the entire installation process (
+ 50 °F
… 104 °F
)! This limit must be adhered to at all times in order to not damage the sensors inside the cable!
Note
A corresponding information is attached to each cable drum or cable ring (document no.
F019
).