36
Clean the Tali
™
Image-Based Cytometer
Introduction
We recommend that the Tali
™
Image-Based Cytometer be cleaned periodically to
prevent the buildup of dust and dirt that might reduce its performance and cause
contamination.
Important
To avoid electrical shock, always turn off the Tali
™
Image-Based Cytometer and
unplug the power cord before cleaning or decontaminating the instrument.
Using a cleaning or decontaminating method other than that specified by the
manufacturer may result in damage to the instrument.
All biological samples and materials that come into contact with them have the
potential to transmit infectious diseases and are considered biohazardous.
Follow all applicable local, state/provincial, and/or national regulations. Wear
appropriate protective eyewear, clothing, and gloves.
Clean the
touch screen
Wipe the touch screen of the Tali
™
Image-Based Cytometer using a soft, lint-free
cloth moistened with an LCD cleaning solution. Do not apply excessive force
during cleaning. Wipe the touch screen dry immediately after cleaning.
Ensure that the cleaning solution does not enter the power button, the power
inlet, the slide port, or the USB ports.
Never pour or spray any liquids directly on the instrument to avoid electrical
shock when the instrument is plugged in.
Do not use abrasive cleaning solutions or material to prevent the touch screen
from getting scratched.
Clean the
instrument case
Wipe the instrument case of the Tali
™
Image-Based Cytometer using a soft, lint-
free cloth moistened with distilled water. Wipe the instrument dry immediately
after cleaning.
Ensure that water or other cleaning solutions do not enter the power button, the
power inlet, the slide port, or the USB ports.
Never pour or spray any liquids directly on the instrument to avoid electrical
shock when the instrument is plugged in.
Decontaminating
the instrument
Wipe the instrument case of the Tali
™
Image-Based Cytometer using a soft, lint-
free cloth moistened with 70% alcohol. Wipe the instrument dry immediately after
cleaning.
Avoid using a bleach solution, because it may leave a residue of bleach crystals on
the instrument. Avoid cleaning the touch screen.
Ensure that water or other cleaning solutions do not enter the power button, the
power inlet, the slide port, or the USB ports.
Never pour or spray any liquids directly on the instrument to avoid electrical
shock when the instrument is plugged in.