Calibrate the Tali™ Image-Based Cytometer,
continued
Calibrate the green
and red channels,
continued
5.
When the green channel calibration is complete, the cellular analysis slide will
automatically be ejected and the Red channel calibration dialog box will be
displayed.
6.
Load 25 μL of Tali
™
Red Calibration Beads into the Tali
™
Cellular Analysis
Slide, insert the slide into the slide port, and touch
Continue
to initiate red
channel calibration.
7.
After the calibration process is completed, touch
OK
to return to the Settings
screen. There is no need to recalibrate each time the instrument is turned on.