Information for the tester
59
Low-Resistance Measurement and Continuity Check (R)
Measuring method:
Current-voltage measurement as per
IEC
61557-4
Nominal voltage:
Battery voltage:
≥
5,5 V; max. 9 V DC
Short-circuit current:
≥
200 mA DC
Permissible overload:
U
rms
= 600 V (measurement is not started)
Measured-value formation
without zero compensation: R
display
= R
measured
- 70 m
Ω
(measuring-lead
resistance)
Measured-value formation
with zero compensation
R
display
= R
measured
- R
compensation
( <o> / section
3.10)
Compensation of measuring cables up to 5
Ω
possible.
Measuring Range
Resolution
Intrinsic error
0,01...0,15...999
Ω
0,01-0,1
Ω
± (5% of m.v.
+ 3 digits)
Internal resistance:
approx. 20
Ω
Measuring time:
approx. 2 s incl. voltage pole change for resistive
specimen
rising to max. 20 s for parallel inductances.
Permissible inductive
component:
max. 2.5 H without influence on measurement result;
max. 5 H without damage
Permissible series-mode
voltage:
≤
3 V AC or DC, no measurement is started beyond
this value
Continuity check:
Same as low-resistance measurement, continuous measurement without pole
change after pressing the key. Response threshold: At Rext < 2
Ω ±
0.5
Ω
audible
signal and „
con
“ is displayed.
Resistance Zero Compensation (<o>)
The measured value stored in this function is automatically subtracted in all
resistance measurements (R
A
, R
S
, R)!!
Measuring Range
Resolution
Intrinsic error
0,01-0,15-2,99-5,0
Ω
0,01-0,1
Ω
± (5% of m.v. + 3 digits)
Measuring time:
approx. 1 sec; no voltage pole change
Nominal voltage:
Battery voltage:
≥
5.5 V; max. 9 V DC
Short-circuit current:
≥
200 mA DC, < 450 mA DC
Summary of Contents for Saturn 100 Plus
Page 2: ......
Page 7: ...Initial power up 5 9 2 1 10 5 4 6 7 8 3 11 A 12...
Page 13: ...Measuring 11 Flowchart of operation...
Page 46: ...Information for the tester 44 4 2 Description of Function...
Page 71: ...Erste Inbetriebnahme 5 9 2 1 10 5 4 6 7 8 3 11 A 12...
Page 76: ...Messfunktionen 10 Flu diagramm f r Bedienung...
Page 107: ...Informationen zum Pr fger t 41 4 2 Funktionsbeschreibung...
Page 126: ...Service 60...
Page 131: ...Premi re mise en service 5 9 2 1 10 5 4 6 7 8 3 11 A 12...
Page 138: ...Directives d utilisation 12 Organigramme d utilisation...
Page 174: ...Informations 48 4 2 Les fonctions...
Page 199: ...Primo Avvio 6 9 2 1 10 5 4 6 7 8 3 11 A 12...
Page 204: ...Funzioni di misura 11 Diagramma di flusso per l utilizzo...
Page 263: ...Puesta en marcha 5 9 2 1 10 5 4 6 7 8 3 11 A 12...
Page 269: ...Medici n 11...