34
Version 1.8, Revision L
Working with the Instrument
5
Key/setting dial
Semiautomatic sectioning mode
Automatic sectioning mode
1
2
–>Rotary knob 2
Specimen retraction (
RETRACT
)
cannot be set.
Note: Value cannot be changed. NO
automatic retraction is possible in
semiautomatic sectioning mode.
–>Rotary knob 2
The specimen retraction (
RETRACT
)
can be set from 0 to 100 μm in 10 μm
increments using rotary knob 1.
–>Rotary knob 2
The LED illumination can be adjusted
to 5 different levels of brightness
using rotary knob 1.
ditto
1
2
–>Rotary knob 2
The feed rate (
FOR/REV
) for the
Blade forward
and the
Blade back
keys can be set between 1 and
5 mm/s in 0.5 mm/s increments with
rotary knob 1.
ditto
1
2
–>Rotary knob 2
Switches the vibration movement
(
FOR/VIB
) of the
Blade forward
key
either on or off using rotary knob 1.
ditto
1
2
–>Rotary knob 2
Key acknowledgment (
BEEP
) on or
off with rotary knob 1.
ditto
Save the parameter and leave by
pressing the
MENU
key.
Note: Leaving the menu and saving
the parameter is possible at any
point in the menu.
ditto
Note
If you want to call up the parameters of a certain user (for example, user 3), press the
MENU
key and
select
user 3
. Then confirm by pressing the
MENU
key 2x. The parameters stored under
user 3
are now
enabled.