45
Leica VT1200 / Leica VT1200 S
Operating the VibroCheck
6
6.
Press the
RUN
key, check the value, repeat steps 5-7 if necessary.
Note
If the display shows
Σμm
0
(optimal) and the value in μm
is unacceptably
high, the blade must be replaced.
7.
If the measured value is accepted, press
DOWN
(flashes green). VC moves to lowest position –
blade is moved to the rear. Control panel of the VT shows:
VIBRO END
. The LED on
the VC is again illuminated in red.
Fig. 30
8.
The instrument now expects the VibroCheck to be removed. To do so, disconnect the USB plug
connection of the VC to the basic instrument and pull the VC off of the dovetail guide. LED in
DOWN
key continues to flash – wait until LED in
DOWN
key goes out.
The normal operating status is now restored.