Malfunctions: Meaning and Troubleshooting
7
Error No. /
INF No.
Error
Troubleshooting
Comment
Error 51
• Horizontal calibration
of VibroCheck is not
possible.
•
Possible operating error,
see the Instructions for
Use.
• Great deal of damage
or dirt on the blade or
blade holder. Use a new
blade or blade holder.
• Controllers have
different software
versions. Have the
software updated.
• VibroCheck is defective.
Have it checked by
service.
• A calibration is carried
out even if the blade is
jagged.
A great deal of obvious
damage or dirt must be
present before a blade
calibration is canceled
automatically.
Error 52
• Basic calibration of light
barrier is not possible
for VibroCheck.
• Emitter diode or receiver
is defective.
• VibroCheck is defective.
• Inform service!
Error 53
• VibroCheck blade
search not successful.
•
No blade and/or blade
holder swung out or not
installed.
• Emitter diode or receiver
dirty. Clean.
• VibroCheck is defective.
• Leica VT1200 drives to
the lowest Z position
using the stepper motor
after an unsuccessful
search. (Allows blade
installation)
Communication
between
keyboard –
Leica VT1200 S
• Communication error
between control unit
and VT
•
Leica VT1200 S display
lights up, but remains
blank.
• Leica VT1200: row of
dots flashes.
• Inform service!
The knife/blade
collides with the
buffer tray
• The buffer tray was not
pushed in up to the rear
pin or was not clamped
down during installation.
•
Push in the buffer tray
as far as it will go and
clamp it down with the
lever.
The VibroCheck
collides with the
blade
• The VibroCheck was not
pushed all the way in or
was not clamped down
during installation.
•
Push in the VibroCheck
as far as it will go and
clamp it down with the
lever.
•
Switching from
SINGLE
to
CONT
is not possible.
• In
MAN
mode, only single cuts are possible.
• The cutting window
cannot be defined.
• In
MAN
mode, only single cuts are possible.
• If multiple cuts are required, switch to
AUTO
mode.