44
Version 1.8, Revision L
Operating the VibroCheck
6
Fig. 26
Fig. 27
3.
The user presses the
DOWN
key. The VC moves to the bottom position, in addition to that the
blade moves to the rearmost position – the LED in the
RUN/STOP
key flashes. Short red flash of
LED on VC --> it then remains red.
4. The user presses the
RUN/STOP
key: First, the blade moves forward (into the position exactly
above the VibroCheck), then the VC moves into a position in which the blade partially covers the
light barrier. LED on VC flashes green –
RUN/STOP
flashes yellow. Control panel shows:
VIBRO
search
see
. A search can last up to 1 minute. When LED on VC is green and LED in
RUN/STOP
is yellow, the blade starts to vibrate.
Fig. 28
1
2
Fig. 29
6
7
8
8
9
The deviation of the height amplitude is shown on the display in μm
. This value
can be either positive or negative. A number (e.g. –0.3)
appears in the Σμm
display. This means a
COUNTERCLOCKWISE
rotation (because of the
-
) by a
turn of 0.3 and reduces the height amplitude to a minimum. (If there is no operation sign, the
rotation is in a clockwise direction
+
.) If the value is
0
, no improvement of the
height amplitude is possible.
5.
Press the
STOP
key. Unscrew the clamping screw
only slightly using the size 3
Allen key, pull the cap
up and off (keep it in a safe place) and turn the adjusting
screw
COUNTERCLOCKWISE
(the
-
direction,
) by a turn of 0.3
using the size 3 Allen key. Tighten the clamping screw