13
Leica VT1200 / Leica VT1200 S
Instrument characteristics
3
Operating noise level
<70 dB
Emissions and Boundary Conditions
Overvoltage category to IEC 61010-1
II
Pollution degree to IEC 61010-1
2
Means of protection to IEC 61010-1
Class I
Degree of protection to IEC 60529
IP20
Heat emission
50J/s
A-weighted noise level, measured at 1 m distance
<70 dB
EMC class Class
B
Electrical Connections and Interfaces
Power supply
Power socket
Electrical Connections
Leica VT1200 S control panel, Leica VibroCheck,
Foot switch
Mechanical Connections
Interfaces to other devices
Magnifier, Microscope, Module LED Hi-Power
spots, 2-arm, Module Hi-Power spot, LED 1000,
Blade holder
Other Specifications
Sectioning frequency (± 10 %)
85 Hz (± 10 %)
Amplitude
from 0 - 3 mm, in 0.05 mm increments
Cutting range
45 mm
Sectioning window
0.5 mm - 45 mm
Specimen orientation, rotating
360 °
Specimen plate, swiveling
0 - 10 °
Electrical overload protection
Yes
Internal current limit for the electronics
Yes
Height with magnifier support
600 mm x 250 mm x 320 mm
Height with microscope
600 mm x 250 mm x 469 mm
L x W x H control unit (when the bases are folded in) 190 mm x 150 mm x 72 mm
Control unit
1 kg
Magnifier support
2 kg
Microscope support with stereomicroscope
4.3 kg
CE
Yes
CSA
Yes
Sectioning speed (± 10 %)
0.01 - 1.5 mm/s
Return speed (± 10 %)
1.0 - 5 mm/s, in increments of 0.5 mm/s
Total vertical specimen stroke
20 mm (motorized)
Specimen retraction
0 - 100 μm (adjustable; can be deactivated)
Maximum specimen size: with standard knife holder
33 x 50 mm
Section thickness selection
manual in 1 μm increments or automatic max.
1000 μm