41
Leica VT1200 / Leica VT1200 S
Operating the VibroCheck
6
6. Operating the VibroCheck
6.1 Using the VibroCheck with the Leica VT1200
Warning
The following instructions must be adhered to exactly. Noncompliance can cause serious damage to
the instrument.
We recommend using the VibroCheck after each blade change to check the optimum position of the blade
and minimize the vertical vibration.
Fig. 20
1
2
3
1
Foot switch for VT
2
VibroCheck
3
Control panel for VT
Fig. 21
4
5
Prior to assembly, the dovetail guide
is brought down to the lowest position using the
DOWN
key!
1.
The user installs the VibroCheck: Push the VC along the dovetail guide
to behind
the mark on the baseplate of the instrument (rear stop) and clamp it down using the lever
. Insert the blade and clamp tightly. Return the blade to the cutting position see
2.
Plug the connecting plug of the VibroCheck (VC) into the socket
panel. Short red flash of LED on VC --> control panel acknowledges VC. LED then remains red.
The LED in the
DOWN
key flashes green.
3.
The user presses the
DOWN
key. The VC moves to the bottom position, after that the blade
moves to the rearmost position – the LED in the
R
UN/STOP
key flashes. Short red flash of LED
on VC --> it then remains red.