14. Maintenance
Fig. 38
14.3 Replacement of the fuses
Turn the instrument off with the mains switch
and pull the mains plug, before replacing the
fuses.
Only use fuses of the same specification! For
the required values, please refer to
chapter 4
- 'Technical data'.
The mains fuses are located in the voltage selector
housing
(3)
above the mains switch at the rear of the
instrument.
Remove the voltage selector housing
(3)
as de-
scribed under 14.2 .
Remove the fuses
(4)
.
Mount new fuses of the same technical specifica-
tion.
The actual voltage setting can be seen in the small win-
dow
(5)
of the voltage selector housing
(3)
.
Insert the voltage selector housing with the fuses
into the instrument, lightly push until it locks in place.
Check if the correct voltage is indicated in window
(5)
.
4
3
4