16
Instruction Manual V1.1 01/2001
7.2 Connection of foot switch and mains
cable
Connecting the foot switch
Check if the foot switch adapter
(1)
is present and
if it is properly connected.
Connect the plug of the foot switch to the foot switch
adapter
(1)
at the rear of the instrument.
Fix the plug by tightening the two screws
(2)
.
The external control panel (optional acces-
sory) can be connected to this socket instead
of the foot switch.
If neither the foot switch nor the external con-
trol panel is connected the instrument will be
inoperational!
Connecting the mains cable
Before connecting the mains cable, check if the
mains switch
(3)
at the rear of the instrument is in
the OFF position ('O').
Check if the supplied mains cable has the appro-
priate plug.
Connect the mains cable to the socket
(4)
.
Exposure to extreme temperature changes
and high air humidity may cause condensa-
tion inside the instrument.
After transporting, please wait at least 2 hours
to allow the instrument to adopt the ambient
temperature before turning it on!
Failure to comply with this may cause dam-
age to the instrument.
7.
Setup
3
4
2
1
2
Fig. 16