30
Instruction Manual V1.1 01/2001
12. Trouble shooting
12.1 Operator and instrument errors
Certain operating and instrument errors are indicated
in the display or by illuminating LEDs of the control
panel.
12.1.1 Error message 'XX.X.'
An additional decimal point appears in the display at
the right. Please contact your technical service cen-
ter.
They will provide information on how to access and
carry out the TEACH-IN program.
12.1.2 Error message '- - -'
The section thickness reading '00.5' is followed by '- -
-' for all the other section thickness settings except
the 0,5 µm setting. This is caused by an operating er-
ror.
All functions of the instrument are inoperational.
Do not press any button.
Turn the instrument off.
Turn the instrument on again following the instruc-
tions given under 7.4 'Power on'.
12.1.3 Error message 'E-1'
This message follows the indication of the software
version after turning the instrument on.
The retraction parameters are lost.
Press any key to continue.
This will set all retraction parameters totheir de-
fault settings.
Sectioning can be continued without problems.
Before continuing, please check if the values that
may have been entered previously are still correct
and change them if required.