20
Instruction Manual V1.1 01/2001
Section counter
The display indication can be converted from the sec-
tion or trimming thickness indication to section count.
Push the µ/n button.
The display reading changes from section thickness
to section count and the lower LED illuminates.
Whenever the section thickness setting is changed,
the section count reading will shortly change over to
the section thickness indication.
Section counter reset
Push CLEAR to reset the section counter.
The section count reading is set to 000.
Note:
The section count will be reset by activation of CLEAR
even if the display indicates the section or trimming
thickness.
Selecting the sectioning mode
Two motorized sectioning modes are available: single
stroke and continuous stroke.
Once the required sectioning mode is selected, the LED
next to the button illuminates.
Sectioning mode 1 - Single stroke
To select, push the button.
When sectioning is started with the foot switch, the
specimen head completes a section and stops in the
upper limit automatically.
Sectioning mode 2 - Continuous stroke
To select, push the button.
When sectioning is started with the foot switch, the
specimen head moves until the process is stopped by
activation of the foot switch.
The specimen head will then stop upon reaching the
next upper limit of the stroke.
8.
Operation
Summary of Contents for RM2155