37
Leica RM 2155 Rotary Microtome
14.2 Changing the setting of the voltage
selector
If the voltage selector is set to an incorrect
voltage, this can cause severe damage to the
instrument!
The voltage selector is set to the voltage that
was indicated on the order form in the fac-
tory. The name plate reads the power rating
of the voltage selector setting.
A change will only be required if the setting
does not correspond to the voltage available
in your laboratory.
If the setting is changed, please make sure to
change the fuses as well!
The actual voltage setting can be seen in the small
window
(5)
of the voltage selector housing
(3)
.
Turn the instrument off with the mains switch
and pull the mains plug before changing the
voltage setting!
Turn the instrument off with the mains switch.
Disconnect the mains plug from the mains power
outlet.
Place the tip of a small screw driver into the small
recess
(1)
of the cover
(2)
and push carefully to re-
move.
Pull out the voltage selector housing
(3)
.
Pull the voltage selector
(6)
out of the housing
(3)
and insert it again so that desired value can be seen
through the small window
(5)
.
Check if the fuses correspond to the new setting of
the voltage selector.
If necessary, replace the fuses by fuses of the re-
quired specification (see 14.3 'Replacement of the
fuses').
Insert the voltage selector housing with the fuses
into the instrument, lightly push until it locks in place.
Check if the correct voltage is indicated in window
(5)
.
14. Maintenance
Fig. 37.2
3
6
Fig. 37.1
5
2
3
1
Summary of Contents for RM2155
Page 2: ......