1. Motorized sectioning is still
continuing.
1. Foot switch was held down to
start sectioning.
2. Foot switch connected without
the foot switch adapter.
3. The RUN/STOP and RUN/
ENABLE buttons of the external
control panel were not release
simultaneously when starting
sectioning.
1. Section thickness potentio-
meter is maladjusted.
1. Rear and front limits not de-
tected.
1. External control panel not
properly connected.
2. Foot switch adapter was not
removed before connecting the
control panel.
1. Stop motorized sectioning by
pressing the foot switch or the
RUN/STOP or RUN/ENABLE
buttons of the control panel.
1. Press the foot switch shortly.
2. Remove foot switch. Connect
the foot switch adapter. Recon-
nect the foot switch.
3. Release both buttons simulta-
neously if possible.
1. Please call the technical
service and ask for instructions
to carry out the TEACH IN
routine.
1. Call for technical service.
1. Connect properly and fix with
the screws provided.
2. Remove foot switch adapter
and connect control panel
directly to the socket of the
microtome.
The coarse feed button allows only
for a specimen movement away
from the knife.
Sectioning motor stops in
sectioning mode 2 (continuous
stroke) before completing the
stroke.
LED reading changes from one
value to another.
This may be accompanied by an
additional decimal point appearing
at the right-hand digit.
HOME and STOP illuminate
simultaneously.
LEDs of the external control panel
inoperational.
12. Trouble shooting
Problem
Possible Cause
Corrective action