9
Leica RM 2155 Rotary Microtome
4. Technical data
Type
RM 2155
RM 2155
RM 2155
RM 2155
Submitted for approval
-
C-UL
VDE
VDE
Nominal voltage
100 V AC ±10%
120 V AC ±10%
230 V AC ±10%
240 V AC ±10%
Nominal frequency
50/60 Hz
50/60 Hz
50/60 Hz
50/60 Hz
Power draw
340 VA
340 VA
340 VA
340 VA
Protective class
I
I
I
I
Mains fuses
2 x T 3.2 A
2 x T 3.2 A
2 x T 3.2 A
2 x T 3.2 A
UL listed
UL listed
UL listed
UL listed
Pollution degree
2
2
2
2
Overvoltage installation category
II
II
II
II
Maximum heat emission
340 J/s
340 J/s
340 J/s
340 J/s
according to IEC-1010, UL 3101, EN 61010
Microtome
Type
RM 2155 Rotary microtome
Section thickness setting
0.5 - 60 µm
from 0.5 to 1 µm in 0.5 µm increment
from 1 to 20 µm in 1 µm increments
from 20 to 60 µm in 5 µm increments
Horizontal specimen feed
27 mm
via step motor
Vertical specimen stroke
70 mm
Max. sectioning area without retraction
65 mm without specimen orientation
Max. sectioning area with retraction
60 mm
Specimen retraction
in motorized operation
10, 30, 70 µm, dependent on sectioning speed
can be deactivated
in manual operation
5, 10, 15, 20 µm
can be deactivated
Repositioning of knife holder base
North/south
±25 mm
East/west
±20 mm
Maximum specimen size (W x H x D)
50 x 60 x 40 mm
Specimen orientation
horizontal
8°
vertical
8°
rotation
±90°
Trimming steps
5, 15, 30 µm
Electric coarse feed
900 µm/s
Sectioning motor:
Sectioning speed
0; 0.5 - 420 mm/s
Return speed
approx. 120 - 420 mm/s
Summary of Contents for RM2155