24
Pad connector underneath the Mech.
The temperature of the Blue Pad is too cold.
Raise the cryostat temperature.
The ECU is malfunctioning.
Call Instrumedics’ 800 customer service
number and arrange for servicing of the ECU
(1-800-237-2772). DO NOT OPEN THE ECU!
♦
The left LED (labeled Pad) always displays a flashing red light.
Cause
Solution
The temperature of the Blue Pad is too warm.
Lower the cryostat temperature.
The ECU is malfunctioning.
Call Instrumedics’ 800 customer service
number and arrange for servicing of the ECU
(1-800-237-2772). DO NOT OPEN THE ECU!
ONLY IF THE OIL BATH OPTION IS INSTALLED:
♦
The right LED (labeled Oil Bath) is always red.
Refer to the
Oil Bath Accessory Kit Troubleshooting Guide
in this manual.
♦
The Flash does not go on when triggered.
Cause
Solution
The ECU is not turned on.
Make sure the ECU is properly connected to an
electrical source. Make sure the On/Off
Switch on the ECU is in the On position
(pushed in). If the LED’s on the front of the
ECU are on, that indicates that the ECU is on.
The ½ inch diameter Flash/Pad Power Cable is
not connected properly.
Make sure the Flash/Pad Power Cable is
properly connected to the ECU.
One or both fuses in the back of the ECU are
blown.
Check the fuses on the back of the ECU and
replace as needed.
NOTE
: If flash unit is
triggered more than 3 times in one minute,
one or both fuses on the back of the ECU will
blow out and will need to be replaced. This is
a protective feature designed to prevent
overheating. Use only 1AMP, SLO-BLO,
Ceramic Core Wound fuses (available from
Instrumedics).
The Flash triggering mechanism is jammed.
Check for ice build-up around the Black Knob
on the Mech and remove if found. If this
problem cannot be corrected, call
Instrumedics’ 800 customer service number