96
Version 1.3, Revision D
Troubleshooting
7
•
Safe reagent filling process fails. Please call service.
Fig. 74
7.3.2 Crash or freeze of master software
Follow below steps to protect the specimens.
1. Take out specimens from the retort.
2. Restart the instrument. If the problem still exists, call Leica Biosystems Technical Service.
7.4
Replacing a fuse
Caution
Before replacing a fuse, switch the instrument off and pull out the power plug.
Use ONLY the fuses with correct value.
Fuse rating: T15 A, 250 V AC
Fuse Order No.: 14049354395
Fig. 75
1
Replacing a fuse
L
A suitable flat-head screwdriver must be used to prevent damage to the fuse holders.
1. Unscrew the two fuse holders
at the rear panel of the instrument with a flat-
head screwdriver and check for damage.
2. Remove the defective fuse from the holder and replace it with a new one.
3. Install the fuse holders back to the rear panel with the flat-head screwdriver.