HistoCore PEARL
Safety
2
2.2.5 Cleaning and maintenance
Caution
•
Do not clean the instrument with solvents containing acetone or xylene. No liquid may be spilled
into the internal components of the instrument – neither during operation nor during cleaning.
•
When using cleaners, please comply with the safety instructions of the manufacturer and the
laboratory safety regulations.
•
Check the condensate bottle at least once a week and empty it.
•
If the instrument is not to be used for a long time, drain out the reagent and paraffin. Use deionized
water to clean the formalin bottle and water bottle, and dry; use ethanol to clean the ethanol
bottle, xylene bottle and paraffin baths, and dry; use ethanol to wipe clean the instrument exterior.
2.3
Safety features on the instrument
The HistoCore PEARL tissue processor is equipped with safety functions and with sophisticated
software control mechanisms. These ensure that the specimens remain undamaged in the event of a
power failure or other malfunction during the program and that the program is completed successfully.
2.3.1 Overpressure protection
•
When the power is switched off, the pressure air pump and valves default to a safe condition (retort
vented, no pressure generation).
•
In addition, there is a relief valve that vents all excess air pump output to atmosphere.
2.3.2 Overcurrent protection
Overcurrent conditions are protected against by the main fuse overcurrent protection.
2.3.3 Overheating protection
The heat limiter protects the instrument against the abnormally high temperature.