![Leica BIOSYSTEMS 14 0493 80101 Instructions For Use Manual Download Page 81](http://html.mh-extra.com/html/leica-biosystems/14-0493-80101/14-0493-80101_instructions-for-use-manual_1873461081.webp)
81
HistoCore PEARL
Operation
5
5. If RMS is set On and Cassette mode is selected, the following box pops up
.
Fig. 66
6. Select the cassette number and press
OK
. The
End Time of Program
for the selected program is shown.
Fig. 67
2
3
4
1
7. Set the day and the time (in 24-hour format)
on which you want the program to
end.
Note
The date sequence is Day-Month-Year.
8. Press
.
Note
If the first reagent in the program belongs to the fixation group, programs can be started up to
7 days in advance. If the program is ended at a defined date/time, the delay step is lengthened
accordingly.
» Or if you do not set the end time, press
ASAP
The program starts. The Dashboard is switched to the program running panel.
Note
After press
ASAP
, the Endtime field
is displayed as
ASAP
, and the
OK
button
changes to
Run
.