![Leica BIOSYSTEMS 14 0493 80101 Instructions For Use Manual Download Page 23](http://html.mh-extra.com/html/leica-biosystems/14-0493-80101/14-0493-80101_instructions-for-use-manual_1873461023.webp)
23
HistoCore PEARL
Initial Operation
4
4.
Initial Operation
4.1
Installation site requirements
•
The instrument requires an installation area of approximately 700 x 800 mm.
•
The substrate must have a sufficient load capacity and rigidity with respect to the weight of the
instrument.
•
The instrument is designed for indoor use only.
•
The instrument must be installed in a place that ensures an easy disconnection from the power
supply. The power cable must be in a place that can be easily reached.
•
No extension cable can be connected.
•
The instrument MUST be connected to a grounded power socket.
•
Use only one of the provided power cables that is intended for the local power supply.
•
Avoid vibrations, direct sunlight and heavy variation in temperature.
Caution
The chemicals used in the HistoCore PEARL are easily inflammable and hazardous to health. Therefore
the installation location must be well ventilated, and must contain no open flames. The room in which the
instrument is located must not be used as a permanent station for people. Otherwise the room must be
equipped with an exhaust device.
The installation location must be protected against electrostatic discharge.
Do not operate the instrument in rooms with explosion hazard.
To ensure proper function of the instrument, it must be set up while maintaining a minimum distance of
10 cm from walls and furniture.
4.2
Standard delivery - packing list
Note
To prevent damage to the instrument or specimens, only accessories and spare parts authorized by
Leica Biosystems can be used.
Quantity
Part description
Order No.
1
HistoCore PEARL instrument
14 0493 50667
1
Display module
14 0493 50642
14
Reagent bottle assembly
14 0493 50465
1
Paraffin drip tray
14 0493 50570
1
Bottom tray
14 0493 55572
1
Bottle label
14 0493 50467
2
Basket assembly
14 0493 51403
1
Paraffin scraper
14 0393 53643
1
Maintenance kit, consists of
14 0493 53391
10 O-rings
14 0493 51361
1 Lubricant, Molykote 111, 100g
14 0336 35460
10 Seal bottle cap, normal
14 0461 36136