HistoCore PEARL
Operation
5
Preventive maintenance reminder
The instrument calculates the operation time , including the idle time. When the time exceeds the
warning threshold, the warning symbol presents in the status bar, and a warning message pops up on
the screen, reminding you to contact your service provider to perform a preventive maintenance.
It can only be reset by the Service User by pressing the
Reset
Caution
Only qualified service personnel authorized by Leica Biosystems can do the yearly preventive
maintenance.
Active carbon filter
The instrument calculates the time of the active carbon filter in use. When the time exceeds the default
warning threshold, a warning pops up, reminding you to replace a new active carbon filter. When the
time exceeds the default alarm threshold, the alarm rings with a warning message.
Reset the active carbon filter
L
The default warning threshold is 45 days. The default alarm threshold is 60 days. Validate your
actual threshold before use.
L
Only Supervisor and Service user level can perform this.
1. Press
Maintenance
2. Press
Reset
.
3. Press
Save
Start Service App
The
Start Service App
button
is for Service User only .