79
HistoCore PEARL
Operation
5
5. Place the filled paraffin bath back in the paraffin oven. When inserting the paraffin bath,
always make sure that it slides back through the sliding slot and connects into the back panel
correctly.
Note
It is recommended to refill the paraffin bath one by one.
6. In the reagent status window, reset the data. For more information, see
Note
If the paraffin baths are refilled with paraffin of different melting points, empty and clean the paraffin
baths and the filter. After refilling, reset the melting point of the paraffin baths, for more information,
see the section of System Setup
.
5.7
Starting a program
Before a program can be started, a number of conditions must be met. The following steps are essential
in order to be able to start a program.
Caution
•
Before switching the instrument on, visually check the instrument and accessories for any
damages possibly caused during transportation or moving the instrument. Do not operate the
instrument and contact your local service representative if you find or suspect any damage.
•
Before processing patient tissue samples after transportation or moving the instrument, perform a test
run to ensure the suitability of the instrument and the program used. Either use a pre-installed program
or combine this activity with the validation of your own protocols. Only process patient tissue samples
for diagnostic purpose, after all above checks have been performed successfully. If any of the previous
steps failed, contact your local service representative.
Caution
The pre-installed programs, the customized processing programs and RMS (Reagent Management
System) have to be validated by the user before running the program, i.e. tissue processing with patient
tissue for diagnostics, according to the local or regional accreditation requirements.
Caution
Check the reagent levels and the paraffin levels before starting a program. To do so, keep your eye-sight
and the reagent/paraffin level at the same level, and make sure that the volume of the reagent/paraffin
is not either more than the max. level or lower than the min. level. The insufficient reagent/paraffin may
cause the specimen to dry out due to the specimen not being covered by sufficient reagents/paraffin.
Check that the reagent bottles and paraffin baths are correctly inserted.
Caution
Never leave a paraffin bath empty. The absence of the paraffin may cause incorrect heating to the paraffin
bath.