24
Version 1.3, Revision D
Initial Operation
4
Quantity
Part description
Order No.
1
Additional reagent bottle assembly
14 0493 53393
1
Funnel
14 0493 55900
2
Connector male 5-pole
14 0493 51429
1
Instructions for Use (printed English
with language CD 14 0493 80200)
14 0493 80001
The country specific power cord needs to be ordered separately. Please find a list of all power cords
available for your device on our website www.LeicaBiosystems.com within the product section.
Note
Please check all delivered parts against the packing list and against your order to verify whether the
delivery is complete! Should you find any discrepancies, please contact your Leica Biosystems sales
office without delay.
4.3
Unpacking and installation
Caution
The packaging has two indicators
that indicate improper transport. When the
instrument is delivered, check this first. If one of the indicators is triggered, the package was not handled
as prescribed. In this case, please mark the shipping documents accordingly and check the shipment for
damage.
Caution
At least TWO persons are required when unpacking the instrument.
If the instrument is moved over stairs, please call your local Leica Biosystems Technical Service, who can
safely transport it with appropriate equipment.
Caution
DO NOT use the handles on the rear panel to carry, as they are not designed to bear the weight, but only
for rolling the instrument.
Unpacking the instrument
L
Before unpacking the instrument, read the unpacking instructions carefully. They are attached to
the outside of the transport packaging.
L
Sufficient surface area is required for unpacking the instrument. The distance to the nearest wall
must be at least 1 m on the side and rear. On the front, the distance must be at least 3 m, as the
HistoCore PEARL is rolled from the pallet in this direction.
L
Bring the instrument package as close as possible to the final installation location.