14. STATUS DISPLAY (STATUS)
174
14.4
Phase Difference Measurement Settings (EXT REF PHASE)
To configure phase difference measurement settings, press F•3 EXT REF PHASE in the
status menu. You can use this menu to measure the phase difference between the SDI signal
and the external sync signal and between channels A and B. You can also measure the
phase difference between links A and B when the link format is set to dual.
STATUS (press and hold)
→
F•3 EXT REF PHASE
→
Figure 14-10 EXT REF PHASE menu
To perform phase difference measurement using an external sync signal, set F•4 EXT REF to
ON, and apply an external sync signal. The supported external sync signal formats are listed
below.
• 1080i/60, 1080i/59.94, and 1080i/50
• 1080p/30, 1080p/29.97, 1080p/25, 1080p/24, and 1080p/23.98
• 1080PsF/30, 1080PsF/29.97, 1080PsF/25, 1080PsF/24, and 1080PsF/23.98
• 720p/60, 720p/59.94, 720p/50, 720p/30, 720p/29.97, 720p/25, 720p/24, and 720p/23.98
• NTSC and NTSC with a 10-field ID
• PAL
[Reference] EXT
REF
→
Section 14.4.6, “Switching to an External Sync Signal”
To measure the phase difference between channels A and B, select Bch using INPUT, and
then set F•3 REF SELECT to Ach.
[Reference] REF
SELECT
→
Section 14.4.4, “Displaying the Phase Difference between Channels A and
B”
To measure the phase difference between links A and B, set the link format to dual, and then
set F•3 REF SELECT to LINK A.
[Reference] Dual
link
→
Section 5.1.1, “Selecting a Link Format”
REF SELECT
→
Section 14.4.5, “Displaying the Phase Difference between Links A and B”