68
Command Type
Description
GO
W
Command to execute current settings.
CN
W
Command to cancel current settings
IE
R
Read current IEEE address
IE=n
W
Set IEEE address to n
HC
R
Read current Height Correction
HC=n
W
Set Height Correction to n
C1
R
Read Pt vacuum calibration point
C1=n
W
Set Pt vacuum calibration point
C2
R
Read Pt full-scale calibration point
C2=n
W
Set Pt full-scale calibration point
C3
R
Read Ps1 and Ps2 vacuum calibration point
C3=n
W
Set Ps1 and Ps2 vacuum calibration point
C4
R
Read Ps1 and Ps2 full-scale calibration point
C4=n
W
Set Ps1 and Ps2 full-scale calibration point
CC
W
Activate new calibration points after new values have
been entered.
SF=0
R
Read the status of the Self Test last performed.
SF=1
W
Start the Self Test.
SG
W
Set 'Ground'
GG
W
Go To 'Ground'
F3
W
Change the main screen to 3-output format
F2
W
Change the main screen to 2-output format