27
3
-output format
Units: knots Feet Feet
Actual: M 101.2
1006 M
1005 M
Target:
20.0
0
Now follow the same process for entering the value as described above in Numeric Value
Entry.
Note: The last effective Ps2 target value is retained through a power On/Off cycle.
Important Note:
If the Ps2 output is not being used then it is mandatory that the Ps2
output be in units of Differential inHg (Dfin) and the Ps2 target value be set to 0.000 Dfin.
This is NOT done automatically on power-On. It is the operator’s responsibility to perform
this function. Failure to do so will result in improper operation of the unit.
Important Note:
If the Ps2 output is not being used then it must be capped-off (closed).
Rate target value :
The Rate target value can only be entered in the 2-output format.
Change the Main Screen to the 2-output format using the PREV SCREEN key. In the 2-output
format the middle vertical section will show the Rate units for Ps1 and will indicate “/min”.
After the screen shows the 2-output format, press the Ps2 / RATE key to enter the Rate target
value. After the key is pressed, the 2-output format will appear as shown below.
2-output format
Units: knots Feet/min
Feet
Actual: M 101.2
3500
1005 M
Target:
20.0
0
Now follow the same process for entering the value as described above in Numeric Value
Entry.
Note that while entering Rate values, it is NOT necessary to include the sign of the rate of
change. The 6600 will automatically determine the direction of the rate of change. The
displayed actual value will, however, indicate with a minus sign when the rate of change is
negative.
Note: The last effective Rate target value is retained through a power On/Off cycle.