35
operation, all three outputs must be connected to the aircraft and all three outputs must be
set to achieve specific target values. If the aircraft requires only the Pt and Ps1 outputs then
the Ps2 output must be capped-off (closed) on the 6600.
Note:
When the 6600 is powered-ON, the Pt mode will always be in MEASURE.
Ps1 Mode :
The Ps1 mode can be selected both in 2-output and 3-output formats. The
Ps1 output has three modes of operation. The three mode are Measure (M), Leak (L) and
Control (C). The power-up mode defaults to Measure (M). To change the mode press the Ps1
key in the “Mode” section of the keypad. The three modes will be displayed on line 4 of the
display, with the cursor at Measure (MEAS), indicated by “MEAS” blinking. The UP
ARROW key can be used for moving the cursor right and the DOWN ARROW key can be
used to move the cursor left. The cursor wraps around the selections, that is, moving the
cursor left past “MEAS” wraps it around to “CTRL”. The sample screen below shows the
three mode choices displayed on line 4 of the display, with the cursor at “MEAS”.
2-output format
Units: knots Feet/min
Feet
Actual: M 101.2
3500
1005 M
Target:
20.0
3000
0
Ps1 mode : MEAS , LEAK, CTRL
Move the cursor to the desired mode and then press ENTER. The selected mode is displayed
on the screen and blinks until the GO key is pressed. The new mode will not become effective
until the GO key is pressed. The sample screen below shows the selected Ps1 mode as
“LEAK”. Notice that the “L” on the right in line 2 is blinking.
2-output format
Units: knots Feet/min
Feet
Actual: M 101.2
3500
1005 L
Target:
20.0
3000
0
Important Note:
When either the Pt, Ps1 or Ps2 Mode is changed to Control (C), ALL
three outputs (Pt, Ps1 and Ps2) are automatically changed to Control. During normal
operation, all three outputs must be connected to the aircraft and all three outputs must be
set to achieve specific target values. If the aircraft requires only the Pt and Ps1 outputs then
the Ps2 output must be capped-off (closed) on the 6600.
Note:
When the 6600 is powered-ON, the Ps1 mode will always be in MEASURE.