3-1-3. Example of Installation Procedure
Th
e installation method diff ers according to the machines used by the customer and the machine
installation conditions, etc. An example of an installation procedure is described below.
Checks and Adjustments before Installation
Be sure to always check the mounting surfaces before installation.
1
Check the scale unit mounting surface
Check the parallelism between the scale unit mounting surface and rotational surface while rotating
the rotating part of the machine.
<Examples of check methods>
• Set a dial test indicator against the scale unit mounting surface and measure.
• Set an optical parallel on the scale unit mounting surface and measure with an auto-collimator.
Scale unit mounting surface
Rotating part
Dial test indicator
2
Adjust the head unit mounting surface
Adjust the clearance between the scale unit mounting surface and the head unit mounting surface.
<Specifi cations>
See section 8 “Specifi cations.”
<Example of check method>
• Use gauges or other tools and measure the distance between the two surfaces at multiple points.
Scale unit mounting surface
<Example of installation using the head unit lower reference surface>
Head unit mounting surface
Head unit mounting base
Gauge
Rotating part
Summary of Contents for SET-HD100
Page 10: ...ii J SET HD100 9 外形寸法図 23 9 1 ヘッドユニット スケールユニット 23 9 2 インターポレータユニット 25 ...
Page 25: ... J 15 SET HD100 梱包袋 キャリーケース 接続ケーブル CS03 05 クッション 4 心出し治具 収納 5 接続ケーブル 収納 心出し治具 ...
Page 26: ...16 J SET HD100 キャリーケース バンド ロックする バンドで固定 6 蓋を閉めてロック 収納完了 7 ...
Page 32: ...22 J SET HD100 8 7 接続ケーブル 項目 仕様 コネクタ 本多通信工業製 PCR E68FA 両端 ケーブル長 5 m ...
Page 35: ... J 25 SET HD100 9 2 インターポレータユニット 110 118 210 298 単位 mm ...
Page 36: ...26 J SET HD100 ...
Page 69: ...SET HD100 E 25 9 2 Interpolator Unit 210 110 118 298 Serial No Unit mm ...
Page 70: ...SET HD100 26 E ...
Page 71: ......