SET-HD100
22
(E)
8-6. Interpolator Unit Specifi cations
Item
Specifi cation
Notes
Operating temperature
range
+10°C to +30°C
No condensation
Storage temperature range
0°C to +50°C
No condensation
Main materials (chassis)
SUS304, etc.
Mass
5 kg or less
Number of sensor input
6 sensors (analog)
Th
e output data from the USB is the angular
data which has been calibrated by 6-sensors.
Total number of
interpolation
2
36
Simultaneous 6-sensors interpolation
(internal)
Number of output bits
2
30
Output from USB
Calibration
Self-calibration
According to Magnescale calibration method
Level error
0.4 Vp-p or less
Speed error
11.4 min
-1
11.4 min
-1
= 11.4 rpm
Power supply voltage
DC 20 to 24 V
Ripple 150 mV or less
Use a DC stabilized power supply or the AC
adaptor BP-24 (sold separately).
Current
5 A (Max. 8 A)
Fuse
Rated voltage : 250 V
Rated current : 15 A
8-7. Connection cable
Item
Specifi cation
Connector
Honda Tsushin Kogyo PCR-E68FA+ (both ends)
Cable length
5 m
Summary of Contents for SET-HD100
Page 10: ...ii J SET HD100 9 外形寸法図 23 9 1 ヘッドユニット スケールユニット 23 9 2 インターポレータユニット 25 ...
Page 25: ... J 15 SET HD100 梱包袋 キャリーケース 接続ケーブル CS03 05 クッション 4 心出し治具 収納 5 接続ケーブル 収納 心出し治具 ...
Page 26: ...16 J SET HD100 キャリーケース バンド ロックする バンドで固定 6 蓋を閉めてロック 収納完了 7 ...
Page 32: ...22 J SET HD100 8 7 接続ケーブル 項目 仕様 コネクタ 本多通信工業製 PCR E68FA 両端 ケーブル長 5 m ...
Page 35: ... J 25 SET HD100 9 2 インターポレータユニット 110 118 210 298 単位 mm ...
Page 36: ...26 J SET HD100 ...
Page 69: ...SET HD100 E 25 9 2 Interpolator Unit 210 110 118 298 Serial No Unit mm ...
Page 70: ...SET HD100 26 E ...
Page 71: ......