SET-HD100
(E)
11
4. Operation
Self-calibration is performed to guarantee the measurement angle. Performing self-calibration enables
compensation of measurement error due to scale unit and head unit installation. Perform self-calibration
aft er mounting the scale unit and head unit to the rotating stage. Also, perform self-calibration regularly
when a factor arises that may aff ect accuracy, such as a change in ambient temperature.
Th
e calibration value data is saved even if the power is turned off . When the power is turned on and
reference point detection is performed, the saved calibration value data can be automatically read and
applied to compensate the measurement angle.
Note
• Check that the installation and connection work is complete before turning on the power.
Operation Flow
Turn on the power.
Check the LEDs.
Perform reference point detection.
(When reference point detection is performed, the initial or previous calibration value data is read.)
Perform calibration.
*
Only when performing self-calibration
4-1. Measurement when Performing Self-calibration
1
Turn on the power
Set the POWER switch on the rear panel of the interpolator unit to ON. Power will be supplied to the
system.
2
Check the LEDs
Check that the POWER, STATUS, CONFIG and SENSOR (1 to 6) LEDs on the front panel of the
interpolator unit light up green.
Note
All the LEDs light up orange immediately aft er the power is turned on. Aft er that, the STATUS LED
turns off before lighting up green.
3
Perform reference point detection
Rotate the rotating stage at a speed of 1 to 10 min
-1
for two full revolutions or more in the clockwise
direction. (See section 9 “Dimensions” for the clockwise direction.)
When the reference point is detected, the MODE LED lights up orange and then changes to lighted up
green.
Note
If the rotating stage is rotated in the counter-clockwise direction (the direction opposite the clockwise
direction), the reference point will not be detected. Be sure to check that the MODE LED changes to
lighted up orange.
Reference
When the reference point detection operation is performed, the initial or previous calibration value
data is read, and that data can be used to output automatically compensated angular data.
Summary of Contents for SET-HD100
Page 10: ...ii J SET HD100 9 外形寸法図 23 9 1 ヘッドユニット スケールユニット 23 9 2 インターポレータユニット 25 ...
Page 25: ... J 15 SET HD100 梱包袋 キャリーケース 接続ケーブル CS03 05 クッション 4 心出し治具 収納 5 接続ケーブル 収納 心出し治具 ...
Page 26: ...16 J SET HD100 キャリーケース バンド ロックする バンドで固定 6 蓋を閉めてロック 収納完了 7 ...
Page 32: ...22 J SET HD100 8 7 接続ケーブル 項目 仕様 コネクタ 本多通信工業製 PCR E68FA 両端 ケーブル長 5 m ...
Page 35: ... J 25 SET HD100 9 2 インターポレータユニット 110 118 210 298 単位 mm ...
Page 36: ...26 J SET HD100 ...
Page 69: ...SET HD100 E 25 9 2 Interpolator Unit 210 110 118 298 Serial No Unit mm ...
Page 70: ...SET HD100 26 E ...
Page 71: ......