SET-HD100
12
(E)
4
Perform calibration
Press the SELF CAL switch.
Rotate the rotating stage at a speed of 1 to 10 min
-1
for fi ve full revolutions or more in the clockwise
direction. (See section 9 “Dimensions” for the clockwise direction.)
Th
e MODE LED changes to blink green. Th
e angular data is loaded and the calibration value data is
calculated.
5
Check that calibration is complete
Rotate the rotating stage at a speed of 1 to 10 min
-1
for fi ve full rotations or more in the clockwise
direction. (See section 9 “Dimensions” for the clockwise direction.)
Check the self-calibration results with the MODE LED.
LED
Self-calibration results
MODE LED lights up green
Self-calibration was successful.
Th
e correct angle compensation value is applied and measurement is possible.
STATUS LED lights up red
Self-calibration failed.
Th
e self-calibration value data prior to pressing the SELF CAL switch will be
applied. Check the installation condition, etc. and perform the procedure again
from Step 1. Be sure to always turn off the power before changing the installation
condition, etc.
If all the LEDs light up green, measurement can be performed.
4-2. Measurement Using the Previous Self-calibration Value Data
When there are no changes in the installation or environmental conditions and the accuracy prior to
calibration is also unchanged, measurement can be performed using the previous calibration value data.
To achieve reliable accuracy, however, it is recommended to perform self-calibration before use.
To perform measurement using the previous calibration value data, perform Steps 1 to 3 of section 4-1
“Measurement when Performing Self-calibration.”
When the reference point detection operation is performed, the previous calibration value data is read, and
that data can be used to output automatically compensated angular data.
Summary of Contents for SET-HD100
Page 10: ...ii J SET HD100 9 外形寸法図 23 9 1 ヘッドユニット スケールユニット 23 9 2 インターポレータユニット 25 ...
Page 25: ... J 15 SET HD100 梱包袋 キャリーケース 接続ケーブル CS03 05 クッション 4 心出し治具 収納 5 接続ケーブル 収納 心出し治具 ...
Page 26: ...16 J SET HD100 キャリーケース バンド ロックする バンドで固定 6 蓋を閉めてロック 収納完了 7 ...
Page 32: ...22 J SET HD100 8 7 接続ケーブル 項目 仕様 コネクタ 本多通信工業製 PCR E68FA 両端 ケーブル長 5 m ...
Page 35: ... J 25 SET HD100 9 2 インターポレータユニット 110 118 210 298 単位 mm ...
Page 36: ...26 J SET HD100 ...
Page 69: ...SET HD100 E 25 9 2 Interpolator Unit 210 110 118 298 Serial No Unit mm ...
Page 70: ...SET HD100 26 E ...
Page 71: ......