13-4
Trigger Method
Model 831 Manual
Level Trigger Method
In the section “Triggers Tab” on page 4-11, it is explained
how the Triggers Menu, shown in Figure 13-2, is used to set
the parameters defining exceedance events.
FIGURE 13-2 Triggers Menu
An exceedance event for which Event History data is
measured and stored is initiated when either of the following
occur:
•
The SPL level (Fast, Slow or Impulse detector) exceeds
the user-defined SPL 1 level
•
The Peak level (Peak detector) exceeds the user-defined
Peak 1 level
When the SPL level also exceeds the SPL 2 level, this is
noted in the Events display of the
Overall
tab, as shown in
FIGURE 5-21 "Overall Tab: Exceedances” on page 5-21.
When the Peak level also exceeds the Peak 2 level or the
Peak 3 level (which should be higher than the Peak 2 level),
this is noted on the
Exceedances
page of the
Overall
tab.
Summary of Contents for 831
Page 1: ...Model 831 Sound Level Meter Manual...
Page 2: ...I831 01 Rev O Manual Larson Davis Model 831 Manual...
Page 123: ...5 44 Any Level Display Model 831 Manual...
Page 135: ...6 12 Run Modes Without Measurement History Model 831 Manual...
Page 155: ...7 20 Recovery After Improper Shutdown Model 831 Manual...
Page 209: ...11 22 Markers Model 831 Manual...
Page 221: ...12 12 Link to Time History Model 831 Manual FIGURE 12 11 Time History Display...
Page 321: ...16 24 Sound Recording Playback Model 831 Manual...
Page 358: ...Model 831 Manual Displays 18 27 FIGURE 18 42 Measurement Displays FIGURE 18 43 Events Displays...
Page 403: ...20 28 Troubleshooting Model 831 Manual...
Page 413: ...21 10 Calibration When The Model 831 Is Locked Model 831 Manual...
Page 421: ...23 4 File System Model 831 Manual...
Page 501: ...A 64 Vibration Sensitivity Model 831 Manual...
Page 513: ...B 12 Sections 5 6 7 and 9 except 9 3 Model 831 Manual...
Page 525: ...C 12 Community Noise Descriptors Model 831 Manual...