Model 831 Manual
Setup Tabs
4-11
Delta
The Delta level is the maximum allowed change in Current
average level (i.e. L
Aeq
) permitted during the time interval
defined below. The minimum level that can be set is 0 and
the maximum is 5.0 dB. The default is 0.2 dB.
Time
The Time is the minimum interval that the measurement
must run. The measurement will run for the interval
specified and then continue until the stability condition is
met. The smallest value that can be entered is 20 seconds
and the largest is 99:59:59 (h:m:s), The default is 20
seconds.
Triggers Tab
Note that the default values for these
parameters are as shown in
FIGURE 4-10.
FIGURE 4-10 Triggers Tab
The
Triggers
tab is used to define trigger levels associated
with exceedance events; instances where the measured
sound level (SPL or Peak) exceeds one of the user-defined
trigger levels. The variation of sound level during a typical
Summary of Contents for 831
Page 1: ...Model 831 Sound Level Meter Manual...
Page 2: ...I831 01 Rev O Manual Larson Davis Model 831 Manual...
Page 123: ...5 44 Any Level Display Model 831 Manual...
Page 135: ...6 12 Run Modes Without Measurement History Model 831 Manual...
Page 155: ...7 20 Recovery After Improper Shutdown Model 831 Manual...
Page 209: ...11 22 Markers Model 831 Manual...
Page 221: ...12 12 Link to Time History Model 831 Manual FIGURE 12 11 Time History Display...
Page 321: ...16 24 Sound Recording Playback Model 831 Manual...
Page 358: ...Model 831 Manual Displays 18 27 FIGURE 18 42 Measurement Displays FIGURE 18 43 Events Displays...
Page 403: ...20 28 Troubleshooting Model 831 Manual...
Page 413: ...21 10 Calibration When The Model 831 Is Locked Model 831 Manual...
Page 421: ...23 4 File System Model 831 Manual...
Page 501: ...A 64 Vibration Sensitivity Model 831 Manual...
Page 513: ...B 12 Sections 5 6 7 and 9 except 9 3 Model 831 Manual...
Page 525: ...C 12 Community Noise Descriptors Model 831 Manual...