Model 831 Manual
Sections 5, 6, 7 and 9 (except 9.3)
B-11
For a temperature change of 5 °C then 30 minutes are required.
For a static pressure change of 5 kPa then 15 seconds are required.
For a humidity change of 30% (non-condensing) then 30 minutes are required.
m) Field Strength > 10 V/m
The Larson Davis model 831 was not measured for field strengths greater than 10 V/m.
n) Radio Frequency Emission
The mode of operation of the 831 that produces the greatest radio frequency emission levels was with
the 831 set to run and with an EXC010 (10' microphone extension cable) used to connect the PRM831
to the 831.
o) AC Power and Radio Frequency Susceptibility
The mode of operation of the 831 that produced the greatest measurement susceptibility to A.C. power
frequency and radio frequency fields was with the 831 set to run, USB cable attached and with an
EXC010 (10' microphone extension cable) between the PRM831 and the 831.
Summary of Contents for 831
Page 1: ...Model 831 Sound Level Meter Manual...
Page 2: ...I831 01 Rev O Manual Larson Davis Model 831 Manual...
Page 123: ...5 44 Any Level Display Model 831 Manual...
Page 135: ...6 12 Run Modes Without Measurement History Model 831 Manual...
Page 155: ...7 20 Recovery After Improper Shutdown Model 831 Manual...
Page 209: ...11 22 Markers Model 831 Manual...
Page 221: ...12 12 Link to Time History Model 831 Manual FIGURE 12 11 Time History Display...
Page 321: ...16 24 Sound Recording Playback Model 831 Manual...
Page 358: ...Model 831 Manual Displays 18 27 FIGURE 18 42 Measurement Displays FIGURE 18 43 Events Displays...
Page 403: ...20 28 Troubleshooting Model 831 Manual...
Page 413: ...21 10 Calibration When The Model 831 Is Locked Model 831 Manual...
Page 421: ...23 4 File System Model 831 Manual...
Page 501: ...A 64 Vibration Sensitivity Model 831 Manual...
Page 513: ...B 12 Sections 5 6 7 and 9 except 9 3 Model 831 Manual...
Page 525: ...C 12 Community Noise Descriptors Model 831 Manual...