Model 831 Manual
Acoustic Calibration
8-11
Outside Range of Normal Sensitivity
When the results of the calibration correspond to a
sensitivity greater than 3 dB outside the range of the nominal
sensitivity for that microphone, the message shown in
FIGURE 8-8 appears.
FIGURE 8-8 Outside Normal Sensitivity Range
When this message is received, it is advisable to scrutinize
the current calibration sensitivity and overload level. Ensure
that the calibration was performed properly and that all of
the equipment is functioning correctly.
> 0.5 dB From Previous Calibration Result
When the results of the calibration indicate a change in
sensitivity greater than 0.5 dB from the previous calibration
results, the message shown in FIGURE 8-9 appears.
FIGURE 8-9 0.5 dB From Previous Calibration
Result
Summary of Contents for 831
Page 1: ...Model 831 Sound Level Meter Manual...
Page 2: ...I831 01 Rev O Manual Larson Davis Model 831 Manual...
Page 123: ...5 44 Any Level Display Model 831 Manual...
Page 135: ...6 12 Run Modes Without Measurement History Model 831 Manual...
Page 155: ...7 20 Recovery After Improper Shutdown Model 831 Manual...
Page 209: ...11 22 Markers Model 831 Manual...
Page 221: ...12 12 Link to Time History Model 831 Manual FIGURE 12 11 Time History Display...
Page 321: ...16 24 Sound Recording Playback Model 831 Manual...
Page 358: ...Model 831 Manual Displays 18 27 FIGURE 18 42 Measurement Displays FIGURE 18 43 Events Displays...
Page 403: ...20 28 Troubleshooting Model 831 Manual...
Page 413: ...21 10 Calibration When The Model 831 Is Locked Model 831 Manual...
Page 421: ...23 4 File System Model 831 Manual...
Page 501: ...A 64 Vibration Sensitivity Model 831 Manual...
Page 513: ...B 12 Sections 5 6 7 and 9 except 9 3 Model 831 Manual...
Page 525: ...C 12 Community Noise Descriptors Model 831 Manual...