A-16
Model 831 Specifications
Model 831 Manual
Min/Max Integration Time
Minimum and maximum integration time for measurement of time-average levels and sound exposure
levels.
DC Output Time Weighting
Follows the SLM Detector: F, S or I
DC Output Voltage Range
0 to +3 Volt (0 to 300 dB)
V0 = SPL/100
SPL = 100*V0
or
Sensitivity = 0.01 V/dB with resolution of 0.001 V
DC Output Impedance
3650
DC Output Recommended Loads
for less than 0.4% error
A 10
DC Voltmeter represents negligible error (-0.036%).
The output resistance can be accounted for in the interface
design. For example, if a chart recorder has an input load of
10 k
the gain can be set to 1.365 to correct for the loading.
Time Average Levels and Sound Exposure Levels, (s)
Minimum
0.1
Maximum
(daily autostore enabled)
Unlimited
Maximum
(daily autostore disabled, errors less
than 0.5 dB)
> 23 days
Dosimeter Metrics: TWA, Dose (s)
Minimum
0.1
Maximum
Unlimited
Table A-13 Min/Max Integration Time
Table A-12 AC/DC Output
Summary of Contents for 831
Page 1: ...Model 831 Sound Level Meter Manual...
Page 2: ...I831 01 Rev O Manual Larson Davis Model 831 Manual...
Page 123: ...5 44 Any Level Display Model 831 Manual...
Page 135: ...6 12 Run Modes Without Measurement History Model 831 Manual...
Page 155: ...7 20 Recovery After Improper Shutdown Model 831 Manual...
Page 209: ...11 22 Markers Model 831 Manual...
Page 221: ...12 12 Link to Time History Model 831 Manual FIGURE 12 11 Time History Display...
Page 321: ...16 24 Sound Recording Playback Model 831 Manual...
Page 358: ...Model 831 Manual Displays 18 27 FIGURE 18 42 Measurement Displays FIGURE 18 43 Events Displays...
Page 403: ...20 28 Troubleshooting Model 831 Manual...
Page 413: ...21 10 Calibration When The Model 831 Is Locked Model 831 Manual...
Page 421: ...23 4 File System Model 831 Manual...
Page 501: ...A 64 Vibration Sensitivity Model 831 Manual...
Page 513: ...B 12 Sections 5 6 7 and 9 except 9 3 Model 831 Manual...
Page 525: ...C 12 Community Noise Descriptors Model 831 Manual...