Timer Operation
NOTE: The timer button allows activation of an interval (countdown) or
stopwatch (elapsed) timer. The timers cannot be operated simultaneously.
To access the main timer menu, press the
Timer
button anytime during normal
operation. The main timer menu is shown on the LCD display. Use the
▲
and
▼
buttons
to highlight the
Interval
or
Stopwatch
Timer. Press the
OK/MUTE
button to select the
highlighted timer function.
Interval Timer Operation
1.
When selected, the Interval Timer menu is displayed on the LCD. The timer
defaults to 0:00:00 (hours:minutes:seconds).
2.
Press and hold the
▲
or
▼
buttons to increase or decrease the timer interval.
3.
When the proper interval is entered on the display, press the
OK/MUTE
button
to start the timer.
4.
When the timer reaches 0:00:00, an audible alarm will sound, and the timer will
reset itself and repeat the countdown.
5.
Press the
OK/MUTE
button to pause the timer.
6.
Press the
Menu
button to clear the interval timer and return to the main timer
menu.
Stopwatch Timer Operation
1.
When selected, the Stopwatch Timer menu is displayed on the LCD. The timer
defaults to 0:00:00.
2.
Press the
OK/MUTE
button to start the timer.
3.
Press the
OK/MUTE
button again to zero the timer.
4.
Press the
Menu
button to clear the stopwatch timer and return to the main
timer menu.
If An Airflow Alert Activates
The most common causes of an Airflow Alert are:
Blockage of the inlet grilles or exhaust outlet.
Removal of the work surface or grille during operation.
Resetting the Airflow Alert System
The Airflow Alert automatically resets to normal operation once the motor speed has
stabilized.