Product Service 1-800-522-7658
18
or
to choose,
OK to select, MENU to escape
Keypad -Relays
Sash sensors
Canopy Alarm
Airflow Sensor
Diagnostics
or
to choose,
OK to select, MENU to return
Current Settings
Restore Default
Diagnostics
Calibration
Configuration
Service
or
to choose,
OK to select, MENU to return
Current Settings
Restore Default
Diagnostics
Calibration
Configuration
Service
or
to choose,
OK to select, MENU to return
Current Settings
Restore Default
Diagnostics
Configuration
Calibration
Service
or
to choose,
OK to select, MENU to return
Current Settings
Restore Default
Calibration
Configuration
Diagnostics
Service
or
to choose,
OK to select, MENU to escape
Canopy Alarm
Sash sensors
Keypad -Relays
Airflow Sensor
Diagnostics
or
to choose,
OK to select, MENU to escape
Airflow Sensor
Sash sensors
Keypad -Relays
Canopy Alarm
Diagnostics
1.
Press the
Menu
button.
2.
Press the
▼
button until the
Services
option is highlighted (it will
turn yellow).
3.
Press
OK/Mute
to enter the Services menu screen.
4.
Press the
▼
button until the
Diagnostic
option is highlighted (it
will turn white).
5.
Press
OK/Mute
to enter the Diagnostic submenu.
6.
Press the
▼
button until the Diagnostic
Airflow Sensor
option is
highlighted (it will turn white). Press
OK/Mute
.
7.
The display shows the status of the airflow sensor. Its current
output, alarm setting and zero point in Millivolts, along with the
corrected Inflow velocity currently being displayed.
Quickstart
How do I set up automatic operation of the blower and lights?
1.
Press the
Menu
button.
MAIN MENU
1. MyLogic
TM
2. Settings
3. Service
or
to choose,
OK to select, MENU to escape
or
to choose,
OK to select, MENU to return
Current Settings
Restore Default
Calibration
Configuration
Diagnostics
Service